Vol. 12, Issue 6, 976-984, June 2002
METHODS
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
DNA microarrays are powerful tools to detect changes in transcript abundance in multiple samples in parallel. However, detection of differential transcript levels requires a reproducible sample (target) preparation method in addition to a high-performance microarray. Therefore, we optimized a target-preparation method that converts the poly(A)+ RNA fraction of total RNA into complementary DNA, then generates biotin-labeled complementary RNA from the cDNA. We measured the efficiency of incorporation of biotin-containing nucleotides by an enzymatic digestion, followed by resolution via analytical high-performance liquid chromatography (HPLC). When the target was hybridized to a sensitive and reproducible microarray platform, low coefficients of variation in both hybridization intensities and differential expression ratios across target preparations were observed. Nearly identical hybridization intensities and expression ratios are observed regardless of whether poly(A)+-enriched RNA or total RNA is used as the starting material. We show the ability to discern biological and production variability through the use of different lots of commercial samples as visualized by hierarchical clustering. Automation of the target-preparation procedure shows equivalence to the manual procedure, reproducible yields of target, and low variability as measured by hybridization to microarrays. Most importantly, RNA mixing experiments show a linear and quantitative amplification in probe hybridization signals for >6000 genes across the entire signal range.
| |
INTRODUCTION |
|---|
|
|
|---|
DNA microarrays provide a powerful means to monitor the relative
transcript abundance of many genes in parallel
(Schena et al. 1995
; Brown and Botstein 1999
; Lockhart and Winzeler
2000
). Several factors are critical to obtain high-quality microarray data, including a reproducible, linear sample (target) preparation method, a sensitive and reproducible microarray platform, accurate image-acquisition software, and efficient, reliable data mining and
computational tools. The ability to measure the variance in each step
of the microarray assay process requires both high-quality manufacturing and a reproducible assay (Ramakrishnan et al. 2002
). The
ability to obtain high-quality data from a microarray platform enables
an investigator to query the performance of the target-preparation procedure and its contribution to the overall variability because even
low levels of variability in the target preparation can now be
uncovered, measured, and tracked.
The available target-preparation methods can be divided into two
groups: first-strand cDNA that is labeled or tagged with a capture
sequence, or the generation of antisense RNA (aRNA) from
double-stranded cDNA during an in vitro transcription (IVT) reaction.
Labeled cDNA can be prepared via direct incorporation of a
fluorophore-labeled nucleotide or through incorporation of an
aminoallyl-labeled nucleotide, followed by coupling to a fluorophore containing an amine-reactive group to the aminoallyl nucleotide (Schena
et al. 1995
; for review, see Lockhart and Winzeler 2000
). Alternatively, the first-strand cDNA can be tagged with a capture sequence that is used for subsequent detection steps (Stears et al.
2000
). DNA microarrays containing short oligonucleotide probes (<35
nucleotides long) require more target for each hybridization, which
requires an amplification method with smaller sample sizes. Typically,
the generation of aRNA (aRNA is also commonly called complementary RNA
or cRNA) is preceded by first-strand synthesis of cDNA using an
oligonucleotide primer containing a bacteriophage T7 RNA polymerase
promoter proximal to an oligo(dT) sequence (van Gelder et al. 1990
;
Eberwine et al. 1992
; Lockhart et al. 1996
). After second-strand cDNA
synthesis and cDNA purification, an IVT reaction is performed using T7
RNA polymerase in the presence of labeled nucleotides. Alternatives to
this labeling strategy produce unlabeled aRNA, followed by a cDNA
synthesis in the presence of a fluorophore-labeled nucleotide (Wang et
al. 2000
). Any target preparation method requires a linear
amplification of the available transcripts to be representative of the
transcript population.
The starting material in microarray experiments can be either
poly(A)+-enriched RNA or total RNA (Lockhart et al. 1996
;
Mahadevappa and Warrington 1999
; Wang et al. 2000
; Yue et al. 2001
).
Use of poly(A)+ RNA ensures a starting sample that is highly
enriched in mRNA and possibly less contaminated by genomic DNA and
total RNA. However, in cases where RNA quantities are limited, such as
biopsy or laser-capture microdissection samples, it may not be possible
to obtain sufficient quantities of total RNA to ensure that adequate
quantities of poly(A)+ RNA can be purified from the total
RNA. Furthermore, the additional handling and manipulation of the
sample during the poly(A)+ RNA enrichment steps may introduce
experimental artifacts. Recently, several reports have detailed methods
and results starting with total RNA (Mahadevappa and Warrington 1999
;
Baugh et al. 2001
; Zarrinkar et al. 2001
). However, none of these
reports has shown the correlation of either expression levels or
expression ratios obtained when starting with poly(A)+ RNA
compared with total RNA.
In this report, we have investigated the reproducibility of a
target-preparation method that copies the messages into double-stranded cDNA, then amplifies the transcripts through the use of an IVT reaction
(van Gelder et al. 1990
; Eberwine et al. 1992
; Lockhart et al. 1996
).
We present data wherein the coefficients of variation in the
hybridization intensities are 6%-9% across target preparations from
an automated target-preparation procedure. This low variability enables
the demonstration of biological variability where it exists and of the
low variability when data generated from poly(A)+-enriched
RNA and total RNA are compared. Lastly, experiments in which total RNA
is mixed from different tissues show that this target-amplification
procedure is linear and authentically represents the starting material.
| |
RESULTS |
|---|
|
|
|---|
Quantitation of Biotinylated Nucleotides into aRNA
The generation of a high and reproducible specific activity of
target aRNA is important for reproducible microarray data with maximal
detection sensitivity. Therefore, the specific activity of the target
sample was measured after optimization of the target-preparation procedure from that reported by others (Langer et al. 1981
; van Gelder
et al. 1990
; Eberwine et al. 1992
; Mahadevappa and Warrington 1999
;
Hill et al. 2000
; Baugh et al. 2001
). Published amplification protocols
can produce a template-independent product in addition to amplifying
the RNA of interest, thereby decreasing the specific activity of the
final aRNA (Baugh et al. 2001
). To ensure that no nonspecific target is
produced using this method, multiple aRNA target preparations showed a
size distribution similar to the cDNA preparation when analyzed by
capillary electrophoresis (data not shown). The amount and uniformity
of biotin incorporation across target preparations were measured by an
analytical method comprised of an enzymatic digestion of the
biotinylated, complex aRNA into mononucleosides, followed by resolution
using HPLC. A typical chromatogram measuring absorbance at 260 nm shows
excellent resolution of the four mononucleosides and the two biotin
nucleoside peaks (Fig. 1). Biotinylated
uridine and biotinylated cytosine were injected separately or
coinjected with the digested aRNA to verify the identity of these peaks
(data not shown). The incorporation frequency of the biotinylated
cytosine and uridine can be calculated using the peak size as measured
at 294 nm and the extinction coefficients of biotinylated uridine
(13,000 M
1 cm
1) and biotinylated cytosine (9300 M
1 cm
1). This calculation shows a fourfold
better incorporation of biotinylated uridine than of the biotinylated
cytidine. Furthermore, when combined with calculations for the
unlabeled nucleosides, these calculations show that a biotinylated
nucleotide was incorporated approximately every eight bases under these
labeling conditions. Using this method, we have found high and
reproducible specific activities across many tissues and target
preparations (data not shown). Furthermore, these data showed the
potential to reduce reagent costs in the assay through elimination of
the biotinylated cytosine, owing to its lower incorporation rate, and
through other measures. Based on experiments when biotinylated cytosine
was eliminated from the in vitro transcription mix, biotinylated
uridine was incorporated approximately every 10 bases, generating a
high specific activity and comparable assay performance (data not
shown). It is important to note that these data are all derived from
digestion of a complex aRNA mixture and might not be representative of
every transcript.
|
A Highly Reproducible Target-Preparation Method
Although a reproducible rate of incorporation of biotin-containing
nucleotides across target preparations is desired, a reproducible transcriptional profile is the true goal. To test the reproducibility of the target-preparation procedure, four different aRNA preparations were generated from the same sample of placental total RNA (commercial source), then hybridized on one slide per preparation. Following the
hybridization, detection, and image analysis, we calculated the
coefficient of variation (CV) for the hybridization intensities for
each probe (Fig. 2). The majority of the
CVs in the hybridization intensities were below 30%, with an average
CV (calculated using every single data point) of 16%. A total RNA
preparation extracted from the human hepatocarcinoma cell line HepG2 by
standard methods using chaotropic salts and phenol was used to generate
two independent target preparations (see Methods). The expression ratio
generated from these two aRNA preparations was plotted as a function of the signal intensity (ideally, every ratio should be equal to 1).
Again, low variability throughout the entire signal range was observed
(Fig. 3A). This variability is less than
the variability reported in Hughes et al. (2001)
. Low variability in
the target preparation should also be evident when differential
expression ratios are examined. To address this issue, expression
ratios were determined by comparing two separate aRNA preparations each for the HepG2 cell line and from liver tissue. A plot of the expression ratio changes in experiment 1 versus those observed in experiment 2 reveals a high correlation (R = 0.984) between the two
experiments (Fig. 3B), as would be expected if there is low variability
in the target-preparation methods.
|
|
A further demonstration of the low target-preparation variability was performed by generating 18 separate aRNA preps over three separate days from the same sample of total RNA across six different tissues or cell lines. The expression levels generated from each of these aRNA preps were then used to cluster the samples according to similarity using a hierarchical clustering algorithm (Fig. 4). We found that the samples clustered into five major clusters and every sample was clustered correctly based on its tissue or cell line of origin. Within the first cluster, there were two clusters corresponding to skeletal and cardiac muscle. Within the second cluster, there were two clusters corresponding to the two different Burkitt's lymphoma cell lines used (Namalwa and CA46V1). Within the liver cluster, there were three clusters corresponding to the three different livers tested from three individuals. The clustering also showed high levels of correlation between the replicate aRNA preps generated from the same RNA sample.
|
Automated Target Preparation
Reproducibility and the number of success rates of the target
preparation could be further enhanced by automation of each reagent
addition, incubations, and purification steps. A QIAGEN robot with
hardware and software modifications to perform all of the processes
required for target preparation was used to generate aRNA (see
Methods). The automated method was found to be robust, with aRNA yields
of 40-120 µg of aRNA generated from 5 µg of total RNA, and highly
reproducible, with CVs in yields ~10% except for lymphoma tissue,
which had a CV of 18% (Fig. 5A).
Significantly, there were zero target-preparation failures (defined as
a preparation with <30 µg of aRNA). This variability in yield
between tissues could be due to the total RNA preparation or to the
percentage of mRNA in each tissue's total RNA pool. Although these
aRNA yields vary slightly less than a nonrobotic aRNA preparation
technique obtained from multiple RNA sources, the observation of
variation is consistent with a published report (Zarrinkar et al.
2001
).
|
The reproducibility of the transcriptional profiles was tested by selecting three random aRNA preparations from each tissue type for hybridization to microarrays. These three aRNA preparations per tissue were hybridized to duplicate microarrays for each target preparation, for a total of six microarrays per tissue. A microarray with sixfold probe redundancy for 1146 probes was used for the hybridizations to reduce the variability introduced by the microarray. For data analysis, the six spots corresponding to each probe were averaged to generate a single value for each probe per microarray to reduce variability attributable to the microarrays. A comparison of the variability versus the signal intensities for four of these tissues shows very low variability for all expression levels (Fig. 5B). In all nine tissues, the average CV for all 1146 probes is below 10% (data not shown).
We next determined whether the automated target preparations produced
equivalent results to manual target preparations. If the methods
produce different microarray results, an investigator would be
precluded from comparing data generated with the two target-preparation
methods. We generated 18 aRNA samples, 6 manually and 12 robotically,
from the same sample of total RNA over 2 d. Each aRNA preparation was
hybridized in duplicate, and Pearson correlation coefficients were
calculated for each pairwise comparison (Table
1). From the data, it is apparent that
there is an excellent linear correlation between hybridization
intensities obtained when aRNA is generated from either the manual or
automated target-preparation method. A comparison of the ratio of the
average intensities for each gene between the manual and robotic data
shows the two methods to be equivalent (average
ratio = 1.07 ± 0.06; maximum ratio = 1.36).
|
Total RNA versus Poly(A)+ RNA
Because the protocol developed in this study starts with total RNA instead of poly(A)+ RNA, we determined whether similar expression levels would be measured regardless of starting material. Poly(A)+ RNA was enriched from large pools of total RNA extracted from human liver or HepG2 cells by standard oligo(dT) chromatography. The total RNA and poly(A)+ RNA samples were used to generate aRNA, then hybridized to microarrays. A more detailed view of the data presented in Figure 4 shows that the signal intensities for each gene are very similar, regardless of whether the starting material was total or poly(A)+ RNA (Fig. 6A). A comparison of the differential expression ratios between the HepG2 and liver aRNA targets shows an excellent correlation between the total and poly(A)+ RNA ratios (Fig. 6B). Significantly, the correlation coefficient is 0.95, and the slope of the regression line is 0.93, indicating very similar transcriptional profiles for aRNA target prepared from total RNA versus poly(A)+ RNA.
|
Sensitivity of Target-Preparation Procedure
The sensitivity of the amplification procedure was measured by adding exogenous spikes of poly(A)-tailed control mRNAs into the input RNA. The results show an equivalent sensitivity of the target-preparation procedure for poly(A)+ RNA versus total RNA when preparing a 1:150,000 or 1:6,000,000 mass:mass dilution, respectively (data not shown). This approximates the percentage of mRNA in total RNA as 2.5%. These comparisons indicate that the sensitivity of the amplification procedure is equivalent from total or poly(A)+ RNA as a function of the percentage of any individual mRNA in the mRNA pool.
A Linear Amplification Procedure
As is evident from these data, low target preparation variability should enable discrimination of samples where more subtle changes in transcript abundance are present. Thus, intensity changes should be reflective solely of the changes in transcript abundance if a linear relationship exists between the hybridization intensities and transcript levels, regardless of whether low or high expressors are examined. We confirmed this scenario by preparing RNA samples that were obtained by mixing brain and Burkitt's lymphoma total RNAs in 20:80, 40:60, 60:40, and 80:20 proportions. Each RNA mixture was used to generate aRNA and hybridized to oligonucleotide arrays. The results show that each gene shows a linear relationship for these RNA mixtures (Fig. 7A,B), although the slope is zero for nondifferentially expressed genes (data not shown). Further examination of the 3592 genes that are differentially expressed more than twofold in this mixture shows a linear response for each gene in the RNA mixtures (data not shown). The linearity of the intensities for each gene indicates that each transcript is amplified in a linear fashion, regardless of whether the gene was expressed at low, medium, or high levels in the two tissues. This same experimental approach was also used with mixtures of brain with placenta and brain with lymphoma RNAs with equivalent linear results (data not shown).
|
An additional question raised is whether the target-amplification and microarray results show the correct fold change for these mixtures. To address this issue, we calculated the ratio of the intensities of individual genes in the 40:20, 80:40, and 80:20 brain:Burkitt's lymphoma mixtures for the 150 highest-expressed genes in brain that are absent in the Burkitt's lymphoma sample. These ratios indicate that, on average, the amplification is not only linear, but also faithfully represents the starting materials. The average amplification corresponds to the ratios of the input RNAs, and the standard deviation is very small (Fig. 7C). The presence of a very different, complex mixture of RNAs and subsequent cDNAs does not appear to affect the amplification procedure of individual probes.
| |
DISCUSSION |
|---|
|
|
|---|
Analysis of differential gene expression using DNA microarrays
requires both accuracy and precision in every step of the manufacturing and assay process. We have presented data here to thoroughly analyze the performance of the target-preparation step of the assay process as
monitored by both an analytical method and by microarray results. We
provide evidence that a commonly used amplification scheme is linear
for thousands of genes (Fig. 7). Additionally, we show that the slope
of the linear amplification depends on the amount of mRNA and that the
average amplification faithfully represents the starting material (Fig.
7C). In other words, a twofold increase in message corresponds to a
twofold increase in signal. These data are in agreement with previous
results showing a linear correlation between in vitro prepared
transcripts spiked into a starting poly(A)+ RNA sample at
prescribed levels (Lockhart et al. 1996
), although the data presented
here measure the naturally occurring transcripts present in total
RNA. Experiments in which we spike in vitro prepared mRNAs into total
RNA have similar results to naturally occurring messages. Furthermore,
we reproducibly detect spiked messages present at 83 parts per billion
for multiple transcripts spiked into total RNA.
This target-preparation method is not only linear and sensitive, it is
also very reproducible. The contribution of the manual target-preparation method to the overall variability is approximately one-third of the total variability, where the additional variability is
introduced by the slide, hybridization, scanning, and spot quantitation
procedures, as calculated using statistical blocking by ANOVA (data not
shown). This conclusion differs from a published report that concluded
that the target preparation had no contribution to the total variance
(Yue et al. 2001
). This discrepancy is most likely due to differences
in the statistical techniques used to identify the sources of
variability. Analysis of the robotic target-preparation method
indicates a lower amount of variability (Fig. 5B). Part of the
target-preparation variance occurs during the preparation of the
reagent mixtures used for the different enzymatic steps (data not
shown). Additional gains in reproducibility will most likely occur with
reduced handling by the end user through the use of dried reagents and
premade reagent mixes.
The robotic target-preparation procedure greatly increases the
throughput of the entire microarray assay. Although the actual amplification technique described here is optimized to reduce the labor
requirements for sample preparation, the throughput for the manual
process is maximally 24 preparations/d per person. The robotic
preparation method increases that throughput to 192 preparations/d,
with less than a 1% failure rate (<30 µg of aRNA produced) when
starting with 5 µg of total RNA input (D. Dorris and F. Dudzik,
unpubl.). The throughput and lower variability of the robotic method
are a useful match for constructing expression profiling databases, for
large experiments, and for molecular classification of cell and tissue
types (DeRisi et al. 1997
; Marton et al. 1998
; Golub et al. 1999
; Iyer
et al. 1999
; Alizedah et al. 2000
; Bittner et al. 2000
; Hughes et al.
2000
; Perou et al. 2000
; Shoemaker et al. 2001
). Identifying molecular
markers for tissue types or disease states requires a stable
target-preparation method as part of an assay that provides sensitivity
and reproducibility, combined with a high-quality microarray
(Ramakrishnan et al. 2002
).
The amount, quality, and type of the starting material are of paramount
importance. Isolation of large quantities of RNA from tissue culture
experiments or large tissue samples permits further purification of a
sample to poly(A)+ RNA. Unfortunately, this type of
experiment will likely be in the minority in the future as expression
profiling moves to biopsy samples, cell-sorted samples, and eventually
laser-capture microdissection samples. Some microarray experiments use
poly(A)+ RNA instead of total RNA because of background and
genomic DNA contamination issues (Stears et al. 2000
). Alternatively,
some methods suggest using >25 µg of total RNA as the starting
material for first-strand cDNA-labeling methods (Yue et al. 2001
). We
can routinely start with <1 µg of total RNA input for the
amplification procedure, followed by a single hybridization (D. Dorris
and D. Trakas, unpubl.). The data presented in Figure 6 indicate that using total RNA as the starting material produces equivalent results to
poly(A)+ RNA. Therefore, using the linear amplification
procedure described here when starting with total RNA is appropriate to
represent the transcriptional profile of the cells being tested. This
detailed comparison of the transcriptional profiles of total RNA versus poly(A)+ RNA indicates that the >95% of the content in
total RNA that is represented by transfer and ribosomal RNAs does not
have a deleterious effect on the amplification of the mRNA when using a
modification of the original approach of Eberwine and colleagues (van
Gelder et al. 1990
; Eberwine et al. 1992
). Whether there is an effect
of tRNA and rRNA on random priming approaches of first-strand cDNA from
total RNA labeling methods is unclear, yet possibly important to fully
define the variability introduced when preparing samples for DNA
microarrays using this method.
| |
METHODS |
|---|
|
|
|---|
RNA Preparation
Total RNA was purchased for all tissues and the Burkitt's lymphoma
cell lines (Ambion; BioChain; Clontech), or prepared from the
hepatocarcinoma cell line HepG2 using the TriZOL method (Invitrogen). The standard TriZOL RNA extraction protocol was modified by adding a
chloroform extraction step following the TriZOL extraction step. HepG2
cells were grown in MEM
plus 10% fetal calf serum in the presence
of penicillin and streptomycin. Poly(A)+ RNA was prepared
using a MEGApure kit according to the manufacturer's instructions (Ambion).
Manual Target Preparation
We added 5 µg of total RNA to a reaction mix in a final volume of
12 µL containing 0.5 pmole/µL T7-(dT)24 oligonucleotide primer. The mixture was incubated at 70°C for 10 min, then chilled on
ice. With the mixture remaining on ice, 4 µL of 5× first-strand buffer, 2 µL of 0.1 M DTT, 1 µL of 10 mM dNTP mix, and 1 µL of Superscript II RNase H
reverse transcriptase (200 U/µL)
were added to a final volume of 20 µL, and the mixture was incubated
in a 42°C waterbath for 1 h. Second-strand cDNA was synthesized in a
final volume of 150 µL, in a mixture containing 30 µL of 5×
second-strand buffer, 3 µL of 10 mM dNTP mix, 4 µL of
Escherichia coli DNA polymerase I (10 U/µL), and 1 µL of
RNase H (2 U/µL) at 16°C for 2 h. The cDNA was purified using a
QIAGEN QIAquick purification kit, dried down by lyophilization, and
resuspended in an IVT reaction mix containing 3.0 µL of nuclease-free
water, 4.0 µL of 10× reaction buffer, 4.0 µL of 75 mM ATP, 4.0 µL of 75 mM GTP, 3.0 µL of 75 mM CTP, 3.0 µL of 75 mM UTP, 7.5 µL of 10 mM Biotin 11-CTP, 7.5 µL of 10 mM Biotin 11-UTP, and 4.0 µL of enzyme mix. The reaction mix was incubated at 37°C for 14 h,
and the aRNA target was purified using an RNeasy Kit (QIAGEN). aRNA
yield was quantitated by measuring the UV absorbance at 260 nm. The
aRNA was fragmented in 40 mM Tris-acetate (TrisOAc; pH 7.9), 100 mM
KOAc, and 31.5 mM MgOAc, at 94°C for 20 min immediately before
hybridization with no further purification steps. This typically
resulted in fragmented target with a size range between 100 and 200 bp.
Target preparations containing spiked transcripts with exogenously
added spikes (prepared in vitro) were added at a mass:mass ratio in a
range from 1:6,000,000 to 1:200,000 into total RNA, or a range
from 1:300,000 to 1:1000 into poly(A)+ RNA.
Automated Target Preparation
The automated target preparations followed the above protocol with the following modifications. All incubations were performed in microtiter plates in a thermal cycler containing a heated lid. All liquid and enzyme solution additions and purifications were performed on a QIAGEN BioRobot (model 9604 or 3000). cDNA purification and aRNA purification were performed using QIAquick 96 PCR purification kits and RNeasy96 kits (QIAGEN), respectively.
Digestion and Chromatography of aRNA
To generate nucleotide monophosphates, 4 U of P1 nuclease were used to digest 20-50 µg of aRNA. The enzyme was incubated with the aRNA at 55°C for 6 h, then at 37°C for 6 h in the presence of 10 U of calf intestine alkaline phosphatase to generate the nucleosides. The digested products were purified using Microcon YM-3 columns followed by centrifugation at 8000g for 30-60 min. The mix was then concentrated using a SpeedVac to 100 µL. This solution was analyzed on an HPLC column equilibrated with 0.03 M TEAA (Solvent A) at a flow rate of 1 mL/min. The following gradient was used: 0%-1% Solvent B (95% AcCN, 5% Solvent A1) over 5 min, 1%-15% Solvent B over the next 15 min, 15%-45% Solvent B over the next 30 min, 45%-100% Solvent B over the next 20 min, and hold at 100% Solvent B for 2 min. The concentration of the heterocycles was determined by the absorbance values at 260 nm (the wavelength where maximal absorption occurs for the heterocycles), and the biotin-containing nucleoside concentrations were determined by the absorbance values at 294 nm (the wavelength where maximal absorption occurs for biotinylated cytosine and biotinylated uridine).
Array Hybridization and Processing
We used 10 µg of fragmented target aRNA for hybridization of each
UniSet Human I Expression Bioarray chip containing 9589 probes (these
9589 probes represent 9203 unique accession numbers [genes],
corresponding to ~8935 unique clusters and 386 control probes,
selected initially from GenBank Unigene build #125) or for
hybridization to a microarray containing probes corresponding to 1100 human genes, each spotted 6 times per array (Motorola Life Sciences).
All probes on these microarrays are 30-mer oligonucleotides spotted by
piezoelectric technologies and covalently attached to a polymeric
surface (Ramakrishnan et al. 2002
). These microarrays were hybridized,
washed, and processed using a direct detection method of the
biotin-containing transcripts by a Streptavidin-Alexa647 conjugate as
described (Ramakrishnan et al. 2002
). Processed slides were scanned
using an Axon GenePix Scanner with the laser set to 635 nm, a PMT
voltage of 600, and a scan resolution of 10 µm.
Data Analysis
Slides were scanned using CodeLink Scanning Software (Motorola Life Sciences), and images for each slide were quantitated using the CodeLink Expression Analysis Software (Motorola Life Sciences). Signal intensities for each spot were calculated by summation of the pixel intensities for each spot, then the local background (based on the median pixel intensity of the area surrounding each spot) was subtracted. Whole array data normalization was performed independently for each slide by dividing each spot's intensity (after background subtraction) by the median signal intensity of all test probes. Hierarchical clustering was performed using the Pearson correlation coefficient in GeneSpring software (Silicon Genetics).
| |
ACKNOWLEDGMENTS |
|---|
We thank Geetha Rajavalu for help with ANOVA blocking.
The publication costs of this article were defrayed in part by payment of page charges. This article must therefore be hereby marked "advertisement" in accordance with 18 USC section 1734 solely to indicate this fact.
| |
FOOTNOTES |
|---|
1 Corresponding author.
E-MAIL David.Dorris{at}emailmot.com; FAX (847) 714-7008.
Article and publication are at http://www.genome.org/cgi/doi/10.1101/gr.227402.
| |
REFERENCES |
|---|
|
|
|---|
Received December 13, 2001; accepted in revised form April 11, 2002.
This article has been cited by other articles:
![]() |
C. Ma, M. Lyons-Weiler, W. Liang, W. LaFramboise, J. R. Gilbertson, M. J. Becich, and F. A. Monzon In Vitro Transcription Amplification and Labeling Methods Contribute to the Variability of Gene Expression Profiling with DNA Microarrays J. Mol. Diagn., May 1, 2006; 8(2): 183 - 192. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-W. Bae, S.-K. Rhee, J. R. Park, W.-H. Chung, Y.-D. Nam, I. Lee, H. Kim, and Y.-H. Park Development and Evaluation of Genome-Probing Microarrays for Monitoring Lactic Acid Bacteria Appl. Envir. Microbiol., December 1, 2005; 71(12): 8825 - 8835. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-W. Bae, S.-K. Rhee, Y.-D. Nam, and Y.-H. Park Generation of subspecies level-specific microbial diagnostic microarrays using genes amplified from subtractive suppression hybridization as microarray probes Nucleic Acids Res., July 19, 2005; 33(13): e113 - e113. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Natsoulis, L. El Ghaoui, G. R.G. Lanckriet, A. M. Tolley, F. Leroy, S. Dunlea, B. P. Eynon, C. I. Pearson, S. Tugendreich, and K. Jarnagin Classification of a large microarray data set: Algorithm comparison and analysis of drug signatures Genome Res., May 1, 2005; 15(5): 724 - 736. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||